The Rat Heparanase (HPA) ELISA Kit is a reliable tool for accurately measuring levels of heparanase in rat serum, plasma, and cell culture supernatants. With high sensitivity and specificity, this kit provides consistent and reproducible results, making it ideal for various research applications.Heparanase is an important enzyme involved in the degradation of heparan sulfate, a key component of the extracellular matrix.
It plays a crucial role in processes such as cell migration, invasion, and angiogenesis, making it a valuable biomarker for studying diseases like cancer, inflammation, and thrombosis.With the Rat Heparanase (HPA) ELISA Kit, researchers can easily quantify heparanase levels in rat samples, allowing for in-depth studies on its function and potential therapeutic applications in various diseases.
Product Name:
Rat Heparanase (HPA) ELISA Kit
Product Code:
RTDL00493
Size:
96 Assays
Target:
HPA
Synonyms:
HPSE, HPR1, HPSE1, HSE1, Endo-glucoronidase
Detection Method:
Sandwich
Range:
15.625-1,000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.