The Rat Cytochrome P450 Reductase (CPR) ELISA Kit is a reliable tool for the quantitative measurement of cytochrome P450 reductase levels in rat samples including serum, plasma, and tissue homogenates. This kit is highly sensitive and specific, providing accurate and reproducible results for various research applications.Cytochrome P450 reductase is an essential enzyme involved in the metabolism of drugs and toxins in the body. Its activity is crucial for the proper function of cytochrome P450 enzymes, which play a key role in drug metabolism, detoxification, and steroid biosynthesis.
Abnormalities in cytochrome P450 reductase expression or activity have been linked to various diseases and drug interactions, making it a valuable biomarker for studying these conditions and developing potential treatments.Overall, the Rat Cytochrome P450 Reductase (CPR) ELISA Kit offers a convenient and efficient method for researchers to study the role of cytochrome P450 reductase in rat physiology and pathophysiology. Its high performance and ease of use make it a valuable tool for advancing scientific understanding in the field of drug metabolism and toxicology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.