The Rat Cluster of Differentiation (CD163) ELISA Kit is specifically developed for the precise measurement of CD163 levels in rat serum, plasma, and tissue homogenates. This kit offers exceptional sensitivity and specificity, guaranteeing consistent and accurate results for various research purposes.CD163 is a vital marker for macrophages and monocytes, playing a key role in regulating inflammatory responses and scavenging hemoglobin-haptoglobin complexes. Its expression levels have been linked to various diseases, including inflammation, diabetes, and cardiovascular disorders, making it a valuable tool for investigating disease mechanisms and potential therapeutic interventions.
With the Rat CD163 ELISA Kit, researchers can efficiently analyze CD163 levels in rat samples, furthering our understanding of immune responses and disease pathogenesis. Order now to accelerate your research and uncover new insights into CD163 biology.
Product Name:
Rat Cluster Of Differentiation (CD163) ELISA Kit
Product Code:
RTDL00190
Size:
96 Assays
Target:
CD163
Synonyms:
M130, MM130, Scavenger receptor cysteine-rich type 1 protein M130, Hemoglobin scavenger receptor
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-99
94
Heparin plasma (n=5)
91-101
96
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.