The Rat Arginase (ARG) ELISA Kit is a reliable and sensitive assay designed for the accurate detection of arginase levels in rat samples. This kit is suitable for use with serum, plasma, and cell culture supernatants, providing researchers with precise and reproducible results for a variety of applications.Arginase is an important enzyme involved in the urea cycle and plays a crucial role in regulating nitric oxide production. Dysregulation of arginase activity has been implicated in various diseases, including cardiovascular disorders, metabolic syndromes, and inflammatory conditions.
By measuring arginase levels, researchers can gain insights into disease mechanisms and potential therapeutic targets.With its high sensitivity and specificity, the Rat Arginase (ARG) ELISA Kit is an invaluable tool for studying arginase biology and its implications in disease pathology. Order now to streamline your research and uncover novel insights into arginase function.
Product Name:
Rat Arginase (Arg) ELISA Kit
Product Code:
RTDL00111
Size:
96 Assays
Target:
Arg
Synonyms:
ARG1, Arginase I, Liver Arginase
Detection Method:
Sandwich
Range:
1.56-100ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
90-101
95
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.