The Rabbit Fibroblast Growth Factor 2 (FGF2) ELISA Kit is specifically designed for the accurate quantification of FGF2 levels in rabbit serum, plasma, and tissue culture supernatants. This kit offers high sensitivity and specificity, ensuring precise and reliable results for a variety of research applications.FGF2, also known as basic fibroblast growth factor, is a key player in cell growth, differentiation, and tissue repair processes. It plays a crucial role in various physiological and pathological conditions, including wound healing, angiogenesis, and cancer development.
The Rabbit FGF2 ELISA Kit provides researchers with a valuable tool for studying the role of FGF2 in these processes and developing potential therapeutic interventions.This ELISA kit is easy to use, with straightforward protocols and quick assay times, making it convenient for high-throughput experiments. With its accurate and reproducible results, the Rabbit FGF2 ELISA Kit is an essential tool for advancing research in the fields of cell biology, regenerative medicine, and cancer biology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.