Mouse Aquaporin 1, Colton Blood Group (AQP1) ELISA Kit
SKU:
MODL00104
UPC:
Product Type:
ELISA Kit
Sensitivity:
0.127ng/mL
Range:
0.312-20ng/mL
ELISA Type:
Sandwich
Synonyms:
AQP-CHIP,CHIP28,CO,Channel-Forming Integral Protein,28kDa,Aquaporin-CHIP,Urine water channel,Water channel protein for red blood cells and kidney proximal tubule
Reactivity:
Mouse
Sample Type:
Serum, Plasma, Tissue Homogenates And Other Biological Fluids
Mouse Aquaporin 1, Colton Blood Group (AQP1) ELISA Kit
The Mouse Aquaporin-1 (Colton Blood Group) ELISA Kit is specifically designed for the accurate measurement of aquaporin-1 levels in mouse serum, plasma, and cell culture supernatants. This kit offers high sensitivity and specificity, ensuring reliable and reproducible results for various research applications.Aquaporin-1 is a key protein involved in regulating water transport across cell membranes, particularly in the kidneys and red blood cells. It plays a crucial role in maintaining water balance and osmotic pressure in the body, making it a valuable biomarker for studying conditions related to fluid regulation and renal function.
With its advanced technology and superior performance, the Mouse Aquaporin-1 ELISA Kit is an essential tool for researchers studying kidney function, red blood cell physiology, and disorders related to fluid balance. Get accurate and dependable results with this reliable assay kit from Assay Genie.
Product Name:
Mouse Aquaporin 1, Colton Blood Group (AQP1) ELISA Kit
Product Code:
MODL00104
Size:
96 Assays
Target:
AQP1
Synonyms:
AQP-CHIP, CHIP28, CO, Channel-Forming Integral Protein,28kDa, Aquaporin-CHIP, Urine water channel, Water channel protein for red blood cells and kidney proximal tubule
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.