The Mouse 5-Hydroxytryptamine Receptor 2C (HTR2C) ELISA Kit is specifically designed for the accurate and precise detection of 5-Hydroxytryptamine Receptor 2C levels in mouse serum, plasma, and tissue homogenates. This kit offers high sensitivity and specificity, ensuring robust and consistent results for a variety of research applications.The 5-Hydroxytryptamine Receptor 2C is a critical receptor involved in serotonin signaling, playing a key role in mood regulation, appetite control, and other physiological functions. Dysregulation of this receptor has been linked to various psychiatric disorders and metabolic conditions, making it an important target for research and drug development.
With its easy-to-use format and reliable performance, the Mouse 5-Hydroxytryptamine Receptor 2C (HTR2C) ELISA Kit is an essential tool for researchers studying serotonin signaling pathways, neurobiology, and related disorders in mouse models. Invest in this kit today for accurate and insightful analysis of 5-Hydroxytryptamine Receptor 2C levels in your experimental samples.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.