Human Interleukin 4 Induced Protein 1 (IL-4I1) ELISA Kit
The Human Interleukin-4 Induced Protein 1 (IL4I1) ELISA Kit is specifically designed for the accurate measurement of IL4I1 levels in human samples such as serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit ensures precise and reliable results, making it perfect for a variety of research applications.IL4I1 is a key protein known for its role in immune responses and inflammatory processes. It is involved in regulating immune cell functions and has been linked to various diseases such as cancer, autoimmune disorders, and infectious diseases.
Studying IL4I1 levels can provide valuable insights into these conditions and aid in the development of potential diagnostic and therapeutic strategies.Overall, the IL4I1 ELISA Kit from Assaygenie is a valuable tool for researchers looking to study the role of IL4I1 in human health and disease, providing accurate and consistent results for their research needs.
Product Name:
Human Interleukin 4 Induced Protein 1 (IL-4I1) ELISA Kit
Product Code:
HUDL01518
Size:
96 Assays
Target:
IL4I1
Synonyms:
FIG1, LAAO, LAO, FIG1, L-amino-acid oxidase
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.