Human Interleukin 18 Binding Protein (IL-18BP) ELISA Kit
The Human Interleukin-18 Binding Protein (IL18BP) ELISA Kit is a powerful tool for the accurate quantification of IL18BP levels in human samples including serum, plasma, and cell culture supernatants. This kit boasts exceptional sensitivity and specificity, ensuring precise and reproducible results for various research applications.IL18BP is a key regulatory protein that binds to and neutralizes interleukin-18, a pro-inflammatory cytokine involved in immune responses and inflammatory processes. By measuring IL18BP levels, researchers can gain valuable insights into the modulation of interleukin-18 activity and its impact on various disease states such as autoimmune disorders, inflammatory conditions, and infectious diseases.
With its high performance and versatility, the Human IL18BP ELISA Kit is an indispensable tool for studying the role of interleukin-18 binding protein in health and disease, providing valuable information for the development of targeted therapies and diagnostic strategies.
Product Name:
Human Interleukin 18 Binding Protein (IL-18BP) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.