The Human Hemoglobin Beta (HBB) ELISA Kit is specifically designed for the precise detection and quantification of human hemoglobin beta levels in serum, plasma, and cell culture supernatants. This kit is highly sensitive and specific, ensuring accurate and repeatable results for a variety of research applications.Hemoglobin beta is a critical component of hemoglobin, the protein responsible for carrying oxygen in red blood cells.
Abnormal levels of hemoglobin beta can indicate various blood disorders and diseases, making it a valuable biomarker for studying these conditions and developing targeted treatments.With its reliable performance and ease of use, the Human HBB ELISA Kit is an indispensable tool for researchers and healthcare professionals seeking to investigate hemoglobin beta levels and their implications in health and disease.
Product Name:
Human Hemoglobin Beta (HBb) ELISA Kit
Product Code:
HUDL01305
Size:
96 Assays
Target:
HBb
Synonyms:
HB-B, HBD, CD113t-C, Beta Globin, Spinorphin
Detection Method:
Competition
Range:
1.56-100?g/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well. And then add 50µL prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.