Human Follicle Stimulating Hormone Receptor (FSHR) ELISA Kit
The Human Follicle Stimulating Hormone Receptor (FSHR) ELISA Kit is specifically designed for the precise measurement of FSHR levels in human serum, plasma, and cell culture supernatants. This kit is highly sensitive and specific, ensuring accurate and reproducible results for a variety of research purposes.The follicle stimulating hormone receptor is essential in the reproductive system, playing a crucial role in regulating follicle development and hormone production in both males and females.
Dysregulation of FSHR has been linked to infertility and other reproductive disorders, making it a valuable biomarker for studying these conditions and exploring potential treatments.This ELISA kit is easy to use and provides quick results, making it a valuable tool for researchers and clinicians working in the fields of reproductive biology, endocrinology, and fertility. Ensure the success of your research with the Human FSHR ELISA Kit.
Product Name:
Human Follicle Stimulating Hormone Receptor (FSHR) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well. And then add 50µL prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.