The Human Folate Receptor 2 (Adult) ELISA Kit is specifically designed for the precise measurement of folate receptor 2 levels in human serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, guaranteeing accurate and consistent results, making it an invaluable tool for various research applications.Folate receptor 2 is a vital protein involved in folate metabolism and transport, playing a crucial role in cell growth and division. Abnormalities in folate receptor 2 expression have been linked to various diseases, including cancer, inflammatory disorders, and neurodevelopmental abnormalities, highlighting its significance as a potential biomarker for disease diagnosis and treatment development.
This ELISA kit provides researchers with a reliable method to quantitatively analyze folate receptor 2 levels in biological samples, facilitating comprehensive investigations into the role of this protein in disease pathology and therapeutic interventions.
Product Name:
Human Folate Receptor 2, Adult (FOLR2) ELISA Kit
Product Code:
HUDL01100
Size:
96 Assays
Target:
FOLR2
Synonyms:
FBP/PL-1, FR-P3, FR-beta, beta-hFR, Folate receptor, fetal/placental, Placental folate-binding protein
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.