Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
The Human Endothelial Cell-Specific Molecule-1 (ESM1) ELISA Kit is a specialized assay developed for the precise measurement of ESM1 levels in human biological samples including serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.ESM1, also known as endocan, is a unique endothelial cell-specific molecule that plays a critical role in inflammation, angiogenesis, and vascular permeability regulation.
Its expression is closely associated with numerous pathological conditions such as cancer, sepsis, and cardiovascular diseases, making it a valuable biomarker for investigating these diseases and potential therapeutic interventions.With its reliable performance and user-friendly format, the Human ESM1 ELISA Kit is an indispensable tool for researchers studying endothelial cell biology, vascular disorders, and related medical conditions. Trust this kit to deliver precise and reliable results for your research needs.
Product Name:
Human Endothelial Cell Specific Molecule 1 (ESM1) ELISA Kit
Product Code:
HUDL00981
Size:
96 Assays
Target:
ESM1
Synonyms:
Endocan
Detection Method:
Sandwich
Range:
15.625-1000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.