The Human Dopachrome Tautomerase (DCT) ELISA Kit is specifically designed for the quantitative detection of DCT levels in human serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, providing accurate and consistent results that are essential for various research applications.Dopachrome tautomerase, also known as DCT, is a key enzyme involved in melanin synthesis and pigmentation. Abnormal levels of DCT have been linked to various skin disorders, including melanoma and vitiligo.
By accurately measuring DCT levels, researchers can gain valuable insights into these conditions and potentially develop innovative treatments.Overall, the Human DCT ELISA Kit is a valuable tool for studying melanin synthesis, skin pigmentation, and related disorders, offering researchers a reliable and efficient method for quantifying DCT levels in biological samples.
Product Name:
Human Dopachrome Tautomerase (DCT) ELISA Kit
Product Code:
HUDL00799
Size:
96 Assays
Target:
DCT
Synonyms:
TYRP2, TRP-2, DT, Dopachrome Delta-Isomerase, Tyrosine-Related Protein 2, L-dopachrome Delta-isomerase
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.