Human Collagen Type XI Alpha 1 (COL11a1) ELISA Kit
The Human Collagen Type XI Alpha 1 (COL11A1) ELISA Kit from AssayGenie is a reliable and accurate tool for detecting levels of COL11A1 in human samples. This kit is designed for use with serum, plasma, and cell culture supernatants, providing researchers with high sensitivity and specificity for reproducible results.Collagen Type XI Alpha 1 is a key component of extracellular matrix structure and is essential for maintaining tissue integrity, particularly in cartilage and connective tissues.
Dysregulation of COL11A1 has been implicated in various diseases, including skeletal dysplasias and osteoarthritis, highlighting its importance as a biomarker for studying these conditions.With this ELISA kit, researchers can confidently measure COL11A1 levels in biological samples, allowing for a better understanding of its role in disease pathology and potential therapeutic interventions. Trust AssayGenie for a reliable solution for your research needs.
Product Name:
Human Collagen Type XI Alpha 1 (COL11a1) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-101
91
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.