Human CD30L (Cluster of Differentiation 30 Lligand) ELISA Kit (HUES02627)
- SKU:
- HUES02627
- Product Type:
- ELISA Kit
- Size:
- 96 Assays
- Uniprot:
- P32971
- Sensitivity:
- 4.69pg/mL
- Range:
- 7.81-500pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Cell Biology
Description
Human CD30L (Cluster of Differentiation 30 Lligand) ELISA Kit
Based on the information in the provided URL, we can create the following product description:The Human CD30L (Cluster of Differentiation 30 Ligand) ELISA Kit from Assaygenie is a reliable and efficient tool for the accurate detection of CD30 Ligand levels in human samples such as serum, plasma, and cell culture supernatants. With high sensitivity and specificity, this kit ensures precise and reproducible results, making it an essential instrument for a variety of research applications.CD30 Ligand, also known as CD153, is a critical protein involved in the regulation of immune responses and cell communication.
It plays a crucial role in conditions such as autoimmune diseases, allergies, and inflammatory disorders, making it a valuable biomarker for studying these conditions and exploring potential therapeutic interventions.Overall, the Human CD30L ELISA Kit from Assaygenie provides researchers with a powerful tool for investigating the role of CD30 Ligand in various diseases and developing new treatment strategies.
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection Method: | Colormetric |
Detection Range: | 7.81-500 pg/mL |
Sensitivity: | 4.69 pg/mL |
Sample Volume Required Per Well: | 100µL |
Sample Type: | Serum, plasma and other biological fluids |
Specificity: | This kit recognizes Human CD30L in samples. No significant cross-reactivity or interference between Human CD30L and analogues was observed. |
This ELISA kit uses Sandwich-ELISA as the method. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CD30L. Standards or samples are added to the appropriate micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CD30L and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CD30L, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by adding Stop Solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human CD30L. The concentration of Human CD30L in samples can be calculated by comparing the OD of the samples to the standard curve.
UniProt Protein Function: | TNFSF8: Cytokine that binds to TNFRSF8/CD30. Induces proliferation of T-cells. Belongs to the tumor necrosis factor family. |
UniProt Protein Details: | Protein type:Membrane protein, integral; Cytokine; Apoptosis Chromosomal Location of Human Ortholog: 9q33 Cellular Component: extracellular space; integral to plasma membrane Molecular Function:cytokine activity; tumor necrosis factor receptor binding; receptor binding Biological Process: cell proliferation; CD8-positive, alpha-beta T cell differentiation; defense response to Gram-positive bacterium; cell-cell signaling; apoptosis; positive regulation of transcription from RNA polymerase II promoter; immune response; signal transduction |
NCBI Summary: | The protein encoded by this gene is a cytokine that belongs to the tumor necrosis factor (TNF) ligand family. This cytokine is a ligand for TNFRSF8/CD30, which is a cell surface antigen and a marker for Hodgkin lymphoma and related hematologic malignancies. The engagement of this cytokine expressed on B cell surface plays an inhibitory role in modulating Ig class switch. This cytokine was shown to enhance cell proliferation of some lymphoma cell lines, while to induce cell death and reduce cell proliferation of other lymphoma cell lines. The pleiotropic biologic activities of this cytokine on different CD30+ lymphoma cell lines may play a pathophysiologic role in Hodgkin's and some non-Hodgkin's lymphomas. Two transcript variants encoding different isoforms have been found for this gene. [provided by RefSeq, Nov 2011] |
UniProt Code: | P32971 |
NCBI GenInfo Identifier: | 416777 |
NCBI Gene ID: | 944 |
NCBI Accession: | P32971. 1 |
UniProt Secondary Accession: | P32971,O43404, |
UniProt Related Accession: | P32971 |
Molecular Weight: | 26,017 Da |
NCBI Full Name: | Tumor necrosis factor ligand superfamily member 8 |
NCBI Synonym Full Names: | tumor necrosis factor (ligand) superfamily, member 8 |
NCBI Official Symbol: | TNFSF8 |
NCBI Official Synonym Symbols: | CD153; CD30L; CD30LG |
NCBI Protein Information: | tumor necrosis factor ligand superfamily member 8; CD30-L; CD30 ligand; CD153 antigen; CD30 antigen ligand |
UniProt Protein Name: | Tumor necrosis factor ligand superfamily member 8 |
UniProt Synonym Protein Names: | CD30 ligand; CD30-L; CD_antigen: CD153 |
UniProt Gene Name: | TNFSF8 |
UniProt Entry Name: | TNFL8_HUMAN |
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | O.D | Average | Corrected |
500 | 2.462 2.488 | 2.475 | 2.404 |
250 | 1.76 1.778 | 1.769 | 1.698 |
125 | 0.957 0.933 | 0.945 | 0.874 |
62.5 | 0.462 0.48 | 0.471 | 0.4 |
31.25 | 0.29 0.274 | 0.282 | 0.211 |
15.63 | 0.179 0.173 | 0.176 | 0.105 |
7.81 | 0.124 0.126 | 0.125 | 0.054 |
0 | 0.061 0.081 | 0.071 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human CD30L were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human CD30L were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 25.01 | 43.42 | 175.58 | 25.58 | 43.99 | 187.42 |
Standard deviation | 1.46 | 2.32 | 8.83 | 1.55 | 2.56 | 6.07 |
C V (%) | 5.84 | 5.34 | 5.03 | 6.06 | 5.82 | 3.24 |
Recovery
The recovery of Human CD30L spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 86-97 | 92 |
EDTA plasma (n=5) | 88-98 | 93 |
Cell culture media (n=5) | 87-100 | 92 |
Linearity
Samples were spiked with high concentrations of Human CD30L and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 86-100 | 86-99 | 90-105 |
Average (%) | 91 | 91 | 98 | |
1:4 | Range (%) | 89-100 | 79-90 | 88-100 |
Average (%) | 95 | 85 | 94 | |
1:8 | Range (%) | 86-98 | 87-99 | 87-98 |
Average (%) | 92 | 93 | 92 | |
1:16 | Range (%) | 90-103 | 84-99 | 83-95 |
Average (%) | 96 | 90 | 88 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro ELISA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent | 1 vial, 10 mL | 4°C(shading light) |
Stop Solution | 1 vial, 10 mL | 4°C |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100µl of standard solutions into the standard wells.
- Add 100µl of Sample / Standard dilution buffer into the control (zero) well.
- Add 100µl of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37°C.
- Aspirate the liquid from each well, do not wash. Immediately add 100µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37°C.
- Aspirate or decant the solution from the plate and add 350µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37°C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 90µL of Substrate Reagent to each well. Cover with a new plate seal and incubate forapproximately 15 min at 37°C. Protect the plate from light. Note: the reaction time can beshortened or extended according to the actual color change, but not by more than 30min.
- Add 50 µL of Stop Solution to each well. Note: Adding the stop solution should be done inthe same order as the substrate solution.
- Determine the optical density (OD value) of each well immediately with a microplate readerset at 450 nm.