Human CCL4L1 (Chemokine C-C-Motif Ligand 4 Like Protein 1) ELISA Kit (HUES01854)
- SKU:
- HUES01854
- Product Type:
- ELISA Kit
- Size:
- 96 Assays
- Uniprot:
- Q8NHW4
- Sensitivity:
- 9.38pg/mL
- Range:
- 15.63-1000pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Cell Biology
Description
Human CCL4L1 (Chemokine C-C-Motif Ligand 4 Like Protein 1) ELISA Kit
The Human CCL4L1 (Chemokine C-C Motif Ligand 4 Like Protein 1) ELISA Kit is a reliable and precise tool for the detection of CCL4L1 levels in human biological samples. This kit offers high sensitivity and specificity, ensuring accurate and reproducible results for a variety of research applications.CCL4L1 is a protein belonging to the chemokine family, known for its role in immune responses and inflammation. It plays a crucial role in recruiting immune cells to sites of infection or injury, influencing the body's immune response. Dysregulation of CCL4L1 has been associated with various diseases, including autoimmune disorders and inflammatory conditions, making it a valuable biomarker for studying these diseases and exploring potential therapeutic interventions.
With this ELISA kit, researchers can confidently measure CCL4L1 levels in serum, plasma, and cell culture supernatants, providing valuable insights into the role of this protein in health and disease. This kit is a valuable tool for researchers seeking to deepen their understanding of immune responses and inflammatory processes, ultimately contributing to the development of new treatments and therapies.
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection Method: | Colormetric |
Detection Range: | 78.13-5000 pg/mL |
Sensitivity: | 46.88 pg/mL |
Sample Volume Required Per Well: | 100µL |
Sample Type: | Serum, plasma and other biological fluids |
Specificity: | This kit recognizes Human CCL4L1 in samples. No significant cross-reactivity or interference between Human CCL4L1 and analogues was observed. |
This ELISA kit uses Sandwich-ELISA as the method. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human CCL4L1. Standards or samples are added to the appropriate micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human CCL4L1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human CCL4L1, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by adding Stop Solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human CCL4L1. The concentration of Human CCL4L1 in samples can be calculated by comparing the OD of the samples to the standard curve.
UniProt Protein Function: | CCL4L1: Chemokine that induces chemotaxis of cells expressing CCR5 or CCR1. Inhibits HIV replication in peripheral blood monocytes that express CCR5. Belongs to the intercrine beta (chemokine CC) family. 10 isoforms of the human protein are produced by alternative splicing. |
UniProt Protein Details: | Protein type:Chemokine; Motility/polarity/chemotaxis; Secreted, signal peptide; Secreted Chromosomal Location of Human Ortholog: 17q12 |
NCBI Summary: | This gene is one of several cytokine genes that are clustered on the q-arm of chromosome 17. Cytokines are a family of secreted proteins that function in inflammatory and immunoregulatory processes. The protein encoded by this family member is similar to the chemokine (C-C motif) ligand 4 product, which inhibits HIV entry by binding to the cellular receptor CCR5. The copy number of this gene varies among individuals, where most individuals have one to five copies. This gene copy contains a non-consensus splice acceptor site at the 3' terminal exon found in other highly similar gene copies, and it thus uses other alternative splice sites for the 3' terminal exon, resulting in multiple transcript variants. [provided by RefSeq, Apr 2014] |
UniProt Code: | Q8NHW4 |
NCBI GenInfo Identifier: | 74727347 |
NCBI Gene ID: | 9560 |
NCBI Accession: | Q8NHW4. 1 |
UniProt Secondary Accession: | Q8NHW4,Q50EM1, Q50EM2, Q50EM3, Q50EM4, Q50EM5, Q50EM6 Q50EM7, Q50EM8, Q569J2, B2RUZ3, B7ZMA8, |
Molecular Weight: | 11,008 Da |
NCBI Full Name: | C-C motif chemokine 4-like |
NCBI Synonym Full Names: | chemokine (C-C motif) ligand 4-like 2 |
NCBI Official Symbol: | CCL4L2 |
NCBI Official Synonym Symbols: | CCL4L; SCYA4L; AT744. 2; SCYQ4L2 |
NCBI Protein Information: | C-C motif chemokine 4-like |
UniProt Protein Name: | C-C motif chemokine 4-like |
UniProt Synonym Protein Names: | Lymphocyte activation gene 1 protein; LAG-1; Macrophage inflammatory protein 1-beta; MIP-1-beta; Monocyte adherence-induced protein 5-alpha; Small-inducible cytokine A4-like |
Protein Family: | C-C motif chemokine |
UniProt Gene Name: | CCL4L1 |
UniProt Entry Name: | CC4L_HUMAN |
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | O.D | Average | Corrected |
5000 | 2.507 2.523 | 2.515 | 2.454 |
2500 | 1.769 1.799 | 1.784 | 1.723 |
1250 | 1.075 1.045 | 1.06 | 0.999 |
625 | 0.502 0.522 | 0.512 | 0.451 |
312.5 | 0.285 0.267 | 0.276 | 0.215 |
156.25 | 0.18 0.164 | 0.172 | 0.111 |
78.13 | 0.115 0.119 | 0.117 | 0.056 |
0 | 0.055 0.067 | 0.061 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human CCL4L1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human CCL4L1 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 230.19 | 696.41 | 2213.50 | 215.95 | 724.84 | 2270.93 |
Standard deviation | 12.06 | 39.07 | 120.19 | 11.81 | 31.31 | 77.89 |
C V (%) | 5.24 | 5.61 | 5.43 | 5.47 | 4.32 | 3.43 |
Recovery
The recovery of Human CCL4L1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 91-105 | 98 |
EDTA plasma (n=5) | 87-98 | 93 |
Cell culture media (n=5) | 85-99 | 92 |
Linearity
Samples were spiked with high concentrations of Human CCL4L1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 90-106 | 85-100 | 92-103 |
Average (%) | 97 | 91 | 97 | |
1:4 | Range (%) | 93-106 | 85-98 | 82-94 |
Average (%) | 98 | 90 | 89 | |
1:8 | Range (%) | 90-105 | 82-93 | 81-92 |
Average (%) | 96 | 87 | 87 | |
1:16 | Range (%) | 91-106 | 81-91 | 80-92 |
Average (%) | 97 | 86 | 87 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro ELISA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent | 1 vial, 10 mL | 4°C(shading light) |
Stop Solution | 1 vial, 10 mL | 4°C |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 90 µL of Substrate Reagent to each well. Cover with a new plate seal and incubate forapproximately 15 min at 37 °C. Protect the plate from light. Note: the reaction time can beshortened or extended according to the actual color change, but not by more than 30min.
- Add 50 µL of Stop Solution to each well. Note: Adding the stop solution should be done inthe same order as the substrate solution.
- Determine the optical density (OD value) of each well immediately with a microplate readerset at 450 nm.