The Human Cat Eye Syndrome Chromosome Region, Candidate 1 (CECR1) ELISA Kit is a highly reliable and accurate tool for the detection of CECR1 levels in human samples such as serum, plasma, and cell culture supernatants. With its high sensitivity and specificity, this kit provides consistent and precise results, making it suitable for a wide range of research applications.CECR1 is a key protein involved in the Cat Eye Syndrome Chromosome Region, Candidate 1, and is known to play a role in various biological processes. By measuring CECR1 levels, researchers can gain insights into its potential role in conditions related to this chromosome region, allowing for a better understanding of its function and potential implications in human health.
Ideal for studying genetic disorders and related conditions, the Human Cat Eye Syndrome Chromosome Region, Candidate 1 (CECR1) ELISA Kit is a valuable tool for researchers looking to explore the role of CECR1 in various diseases and disorders. With its precision and reliability, this kit is essential for advancing our knowledge of genetic pathways and their impact on human health.
IDGFL, ADGF, ADA2, Cat Eye Syndrome Critical Region Protein 1
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.