Human Acetyl Coenzyme A Carboxylase Alpha (ACACa) ELISA Kit
The Human Acetyl-Coenzyme A Carboxylase Alpha (ACACA) ELISA Kit is specifically designed for the precise detection of ACACA levels in human serum, plasma, and cell culture supernatants. This kit is known for its high sensitivity and specificity, ensuring consistent and dependable results for a variety of research applications.ACACA is an important enzyme involved in fatty acid synthesis and regulation of metabolism. Dysregulation of ACACA activity has been linked to various metabolic disorders, making it a valuable biomarker for studying these conditions and developing targeted treatments.
The Human ACACA ELISA Kit from Assay Genie provides researchers with a reliable tool for quantifying ACACA levels, allowing for a better understanding of its role in metabolic processes and potential implications for disease management. Upgrade your research capabilities with this advanced ELISA kit from Assay Genie.
Product Name:
Human Acetyl Coenzyme A Carboxylase Alpha (ACACa) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.