The General Cortisone (COR) ELISA Kit is a cutting-edge tool designed to accurately measure levels of cortisone in various biological samples, including serum, plasma, and tissue culture supernatants. With its high sensitivity and specificity, this kit provides reliable and reproducible results, making it invaluable for a wide range of research applications.Cortisone, a naturally occurring hormone in the body, plays a vital role in regulating metabolism, immune response, and stress levels.
Monitoring cortisone levels can provide valuable insights into various health conditions, including adrenal dysfunction, inflammation, and autoimmune disorders. This kit is essential for researchers and healthcare professionals studying these conditions and developing potential therapeutic interventions.Order the General Cortisone (COR) ELISA Kit today and unlock the potential for groundbreaking research in the field of endocrinology and hormone regulation.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
88-102
95
EDTA plasma (n=5)
90-98
94
Heparin plasma (n=5)
90-102
96
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well. And then add 50µL prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.