The Chicken Glucagon GC ELISA Kit is specifically designed for the precise measurement of glucagon levels in chicken serum, plasma, and tissue lysate samples. This kit offers high sensitivity and specificity, providing accurate and consistent results for a variety of research applications.Glucagon is a vital hormone that plays a key role in regulating blood sugar levels by stimulating the liver to release glucose into the bloodstream. It is essential for maintaining glucose homeostasis and is often dysregulated in conditions such as diabetes and metabolic disorders.
By accurately measuring glucagon levels, researchers can gain valuable insights into the role of this hormone in various physiological processes and diseases, potentially leading to the development of novel therapeutic strategies. The Chicken Glucagon GC ELISA Kit is a valuable tool for studying glucagon biology and its implications for health and disease.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well. And then add 50µL prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.