The Bovine Glutamate Dehydrogenase 1 (GDH) ELISA Kit from AssayGenie is specifically designed for the accurate detection of bovine GDH levels in serum, plasma, and cell culture supernatants. This kit offers high sensitivity and specificity, ensuring reliable and reproducible results for a variety of research applications.Glutamate dehydrogenase (GDH) is a vital enzyme involved in the metabolism of glutamate, playing a key role in energy production and amino acid synthesis.
Abnormalities in GDH levels have been linked to various diseases such as liver damage, neurological disorders, and metabolic disorders, making it a valuable biomarker for studying these conditions and potential therapeutic interventions.With its ease of use and precise results, the Bovine Glutamate Dehydrogenase 1 (GDH) ELISA Kit is an essential tool for researchers working in the fields of biochemistry, molecular biology, and biomedical sciences.
Product Name:
Bovine Glutamate Dehydrogenase 1 (GDH) ELISA Kit
Product Code:
BODL00060
Size:
96 Assays
Target:
GDH
Synonyms:
GLUD1, Glutamate dehydrogenase 1, mitochondrial
Detection Method:
Sandwich
Range:
7.8-500U/L
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.